50ps_GOI_GATE_PROFILE
Ultra Fast Profile

See how we measure the optical gain profile

Typical utra-fast 11mm GOI gate width at fastest setting.
The measured gate width (65ps) is a convolution of the
true gate width and the 40ps width of the laser diode used for the test.
The deconvolved gate width is 50ps.